Comparison of Central Corneal Thickness with Ultrasound Pachymetry, Noncontact Specular Microscopy and Spectral Domain Optical Coherence Tomography
- Year
- 2018
- Type
- article
Data source split
- YÖKSİS YÖKSİS article record
- YÖKSİS venue Seminars in Ophthalmology
- Catalog match (ISSN) Seminars in Ophthalmology
- OpenAlex OpenAlex enrichment (abstract, citations, topics)
Abstract
OpenAlex · English
BACKGROUND: To determine the agreement of central corneal thickness (CCT) measurements taken with ultrasonic pachymetry (USP), spectral domain optical coherence tomography (SD-OCT) and noncontact specular microscopy (NSM). METHODS: A prospective, observational, cross-sectional study was performed in the outpatient ophthalmology clinic. CCT was measured in a total of 147 eyes of 147 consecutive healthy patients with USP, NSM, and SD-OCT. Same examiner performed all examinations. Bland-Altman plots were used to evaluate the agreement between instruments. RESULTS: The average CCT values obtained by USP, NSM, and SD-OCT were 555 ± 37 µm, 554 ± 34 µm, and 546 ± 34 µm, respectively. There was a strong correlation between instruments: USP with SD-OCT (r = 0.937, p < 0.01), USP with NSM (r = 0.943, p < 0.01) SD-OCT with NSM (r = 0.975, p < 0.01) for CCT. The mean differences (lower/upper limit of agreement) for CCT measurements were -10 ± 12.9 µm (15.28/-35.28) between SD-OCT and USP, -8.1 ± 7.7 µm (7/-23.2) between SD-OCT and NSM, and 1.8 ± 12.3 µm (25.9/-22.3) between USP and NSM. CONCLUSIONS: USP and NSM were found to have comparable CCT measurements and these two methods can be used correspondingly. However, CCT measurements by SD-OCT were lower when compared to other methods.
Topics
Citations
OpenAlex cited_by_count. Not a WoS or Scopus citation count; those sources have no separate column here.
15 citations
OpenAlex cited_by_count (cache / database)
3 publications in the local catalog that cite this work (OpenAlex reference match; not the full global list).
- Central Corneal Thickness Measurements With Different Imaging Devices: Ultrasound Pachymetry, Noncontact Specular Microscopy and Tono-pachymetry 2022
- Comparison of Six Methods of Central Corneal Thickness Measurement in Healthy Eyes 2021
- Comparison of central corneal thickness measurements of five different devices with ultrasound pachymetry in healthy eyes 2021