Comparison of central corneal thickness measurements of five different devices with ultrasound pachymetry in healthy eyes

ISSN2459-1777
YÖKSİS OpenAlex Open access · diamond SJR Q3 JCR Q4 TR Index
Year2021
Citations3OpenAlex
Percentile%51.5
FWCI0.261.00 = world average
Scopus (SJR)Q3
WoS (JCR)Q4

Data source split

  • YÖKSİSYÖKSİS article record
  • YÖKSİS venueBeyoglu Eye Journal
  • Catalog match (ISSN)Beyoglu Eye Journal
  • OpenAlexOpenAlex enrichment (abstract, citations, topics)

Abstract

OpenAlex English

OBJECTIVES: Several methods of measuring central corneal thickness (CCT) have been compared, however, the data are still limited and conflicting. The aim of this study was to determine the agreement of CCT measurements performed in healthy eyes using ultrasound pachymetry (USP), non-contact tono/pachymetry, specular microscopy, biometry, Scheimpflug-based corneal topography, and optical coherence tomography (OCT). METHODS: All of the participants underwent a complete ophthalmological examination. The CCT of all of the eyes included was measured using 6 different methods. The agreement between the methods was analyzed using the mean difference and Bland-Altman analysis based on a 95% limits of agreement. RESULTS: A total of 64 patients with a mean age of 40.96±14.52 years (range: 20-78 years) were included in the study. The mean CCT value was 552.10±35.65 µm, 550.40±35.55 µm, 554.67±35.49 µm, 545.39±34.21 µm, 546.25±35.49 µm, and 552.64±33.59 µm using USP, non-contact tono/pachymetry, non-contact specular microscopy, biometry, Scheimpflug-based corneal topography, and OCT, respectively. The bias values determined by Bland-Altman plots were -1.70, -2.56, 6.71, 5.85, and -0.53 for tono/pachymetry, specular microscopy, biometry, topography, and OCT, respectively. OCT demonstrated the lowest bias compared to USP. The overall intraclass correlation coefficient was 0.961 (range: 0.945-0.974) with a 95% confidence interval. CONCLUSION: All of the CCT measurements obtained using non-contact tono/pachymetry, non-contact specular microscopy, biometry, Scheimpflug-based corneal topography, and OCT were consistent with the USP measurements of healthy controls. Larger prospective studies to determine the interchangeability of different methods for CCT measurements in pathological conditions are warranted.

Topics

Citations

OpenAlex cited_by_count. Not a WoS or Scopus citation count; those sources have no separate column here.

3citationsOpenAlex · cited_by_count (cache / database)

Authors

7
  1. TURGAY UÇAK 1
  2. EREL İÇEL 2
  3. NURDAN GAMZE TAŞLI 3
  4. YÜCEL KARAKURT ERZİNCAN BİNALİ YILDIRIM ÜNİVERSİTESİ 4
  5. HAYATİ YILMAZ 5
  6. ADEM UĞURLU 6
  7. MEHMET DEMİR 7