RF safety of wires in interventional MRI: Using a safety index
Journal
Magnetic Resonance in Medicine- Year
- 2001
- Type
- article
Data source split
- YÖKSİS venue Magnetic Resonance in Medicine
- OpenAlex OpenAlex enrichment (abstract, citations, topics)
Abstract
OpenAlex · English
With the rapid growth of interventional MRI, radiofrequency (RF) heating at the tips of guidewires, catheters, and other wire-shaped devices has become an important safety issue. Previous studies have identified some of the variables that affect the relative magnitude of this heating but none could predict the absolute amount of heating to formulate safety margins. This study presents the first theoretical model of wire tip heating that can accurately predict its absolute value, assuming a straight wire, a homogeneous RF coil, and a wire that does not extend out of the tissue. The local specific absorption rate (SAR) amplification from induced currents on insulated and bare wires was calculated using the method of moments. This SAR gain was combined with a semianalytic solution to the bioheat transfer equation to generate a safety index. The safety index ( degrees C/(W/kg)) is a measure of the in vivo temperature change that can occur with the wire in place, normalized to the SAR of the pulse sequence. This index can be used to set limits on the spatial peak SAR of pulse sequences that are used with the interventional wire. For the case of a straight resonant wire in a tissue with very low perfusion, only about 100 mW/kg/ degrees C spatial peak SAR may be used at 1.5 T. But for < or =10-cm wires with an insulation thickness > or =30% of the wire radius that are placed in well-perfused tissues, normal operating conditions of 4 W/kg spatial peak SAR are possible at 1.5 T. Further model development to include the influence of inhomogeneous RF, curved wires, and wires that extend out of the sample are required to generate safety indices that are applicable to common clinical situations. We propose a simple way to ensure safety when using an interventional wire: set a limit on the SAR of allowable pulse sequences that is a factor of a safety index below the tolerable temperature increase.
Topics
Citations
OpenAlex cited_by_count. Not a WoS or Scopus citation count; those sources have no separate column here.
225 citations
OpenAlex cited_by_count (cache / database)
45 publications in the local catalog that cite this work (OpenAlex reference match; not the full global list).
- RF heating due to conductive wires during MRI depends on the phase distribution of the transmit field 2002
- Reduction of implant RF heating through modification of transmit coil electric field 2010
- Multifunctional interventional devices for MRI: A combined electrophysiology/MRI catheter 2002
- MRI‐guided coronary catheterization and PTCA: A feasibility study on a dog model 2003
- Real Time Magnetic Resonance Imaging Guided Endovascular Recanalization of Chronic Total Arterial Occlusion in a Swine Model 2006
- Reduction of the radiofrequency heating of metallic devices using a dual-drive birdcage coil 2013
- Advances in Radiologic Evaluation of Small Bowel Diseases 2013
- Reduction of radiofrequency heating of metallic devices using a dual drive birdcage coil 2013
- Invasive human magnetic resonance imaging Feasibility during revascularization in a combined XMR suite 2005
- Magnetic resonance image‐guided trans‐septal puncture in a swine heart 2005