Article detail · 2018 · article
A temperature sensor implant for active implantable medical devices for in vivo subacute heating tests under MRI
Data source split
- YÖKSİSYÖKSİS article record
- YÖKSİS venueMAGNETIC RESONANCE IN MEDICINE
- Catalog match (ISSN)Magnetic Resonance in Medicine
- OpenAlexOpenAlex enrichment (abstract, citations, topics)
Abstract
PURPOSE: To introduce a temperature sensor implant (TSI) that mimics an active implantable medical device (AIMD) for animal testing of MRI heating. Computer simulations and phantom experiments poorly represent potential temperature increases. Animal experiments could be a better model, but heating experiments conducted immediately after the surgery suffer from alterations of the thermoregulatory and tissue properties during acute testing conditions. Therefore, the aim of this study was to introduce a temperature sensor implant that mimics an AIMD and capable of measuring the electrode temperature after implantation of the device without any further intervention at any time after the surgery in an animal model. METHODS: A battery-operated TSI, which resembled an AIMD, was used to measure the lead temperature and impedance and the case temperature. The measured values were transmitted to an external computer via a low-power Bluetooth communication protocol. In addition to validation experiments on the phantom, a sheep experiment was conducted to test the feasibility of the system in subacute conditions. RESULTS: The measurements had a maximum of 0.5°C difference compared to fiber-optic temperature probes. In vivo animal experiments demonstrated feasibility of the system. CONCLUSION: An active implant, which can measure its own temperature, was proposed to investigate implant heating during MRI examinations. Magn Reson Med 79:2824-2832, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Topics
Citations
OpenAlex cited_by_count. Not a WoS or Scopus citation count; those sources have no separate column here.
27citationsOpenAlex · cited_by_count (cache / database)
8 publications in the local catalog that cite this work (OpenAlex reference match; not the full global list).
- 2020 A Review of Bioresorbable Implantable Medical Devices: Materials, Fabrication, and ImplementationCitations 150 · OpenAlex
- 2020 MR safety watchdog for active catheters: Wireless impedance control with real‐time feedbackCitations 22 · OpenAlex
- 2022 A workflow for predicting temperature increase at the electrical contacts of deep brain stimulation electrodes undergoing MRICitations 17 · OpenAlex
- 2021 MRI Compatible Fiber Optic Multi Sensor Platform for Real Time Vital MonitoringCitations 16 · OpenAlex
- 2021 MRI Compatible Fiber Optic Multi Sensor Platform for Real Time Vital MonitoringCitations 16 · OpenAlex
- 2021 MRI Compatible Fiber Optic Multi Sensor Platform for Real Time Vital MonitoringCitations 16 · OpenAlex
- 2019 Wireless control of induced radiofrequency currents in active implantable medical devices during MRICitations 12 · OpenAlex
- 2025 Passive ultrasonic communication link for deep-tissue sensor implantsCitations 6 · OpenAlex