Comparison of results of two hematological analyzer systems: Dirui BF-7200 and Sysmex XN-1000

ISSN1303-829X
YÖKSİS OpenAlex Open access · gold TR Index
Year2024
Citations2OpenAlex
Percentile%56.0
FWCI0.281.00 = world average
Scopus (SJR)Q3
WoS (JCR)Q4

Data source split

  • YÖKSİSYÖKSİS article record
  • YÖKSİS venueTurkish Journal of Biochemistry
  • Catalog match (ISSN)Turkish Journal of Biochemistry
  • OpenAlexOpenAlex enrichment (abstract, citations, topics)
  • Semantic Scholarcitation count (not merged with OpenAlex)

Abstract

OpenAlex English

Abstract Objectives Complete blood count (CBC) is performed using automated hematology analyzers. It is important that CBC results are comparable, reproducible, and reliable. In this study, our aim is to compare the results of Sysmex XN-1000 and Dirui BF-7200 hematology analyzers. Methods Patient samples randomly selected from the routine workflow for each instrument were measured 20 consecutive times to assess reproducibility. The mean, standard deviation, and coefficient of variation (CV%) were calculated for each hematological parameter. A comparison of results from the evaluated Dirui BF-7200 system with those from the current hematology analyzer Sysmex XN-1000 system was made for all of the samples included in the study. The compatibility between the parameters was evaluated using Passing–Bablok and Bland–Altman analyses. Results The within-run CV% values were outside the desirable biological variation database (European Federation of Clinical Chemistry and Laboratory Medicine) specification for CV% for eosinophil, basophil, mean corpuscular hemoglobin, mean corpuscular hemoglobin concentration, mean platelet volume and platelet distribution width expressed as standard deviation in the Sysmex XN-1000 instrument and eosinophil, basophil, hematocrit and indexes of red blood cell and platelet in the Dirui BF-7200 instrument. When the Bland–Altman and Passing–Bablok analysis results were evaluated together, most parameters showed poor agreement; only white blood cells and lymphocytes showed good agreement between the two instruments. Conclusions As there is variability between results from different hematology analyzers, we recommend analyzing patient samples in the same laboratory using the same analyzer to avoid different results that could be misinterpreted.

Topics

Citations

OpenAlex cited_by_count. Not a WoS or Scopus citation count; those sources have no separate column here.

2citationsOpenAlex · cited_by_count (cache / database)

1 publications in the local catalog that cite this work (OpenAlex reference match; not the full global list).

  1. 2024 A brief history of hematology analyzers and recent advancements: the available testing wealthCitations 1 · OpenAlex

Authors

4
  1. MUHAMMED SEYİTHANOĞLU KAHRAMANMARAŞ SÜTÇÜ İMAM ÜNİVERSİTESİ 1
  2. HİLMİ İSMET TUNCER 2
  3. FATMA İNANÇ TOLUN 3
  4. METİN KILINÇ 4