Skip to content
akaturk Academic measurement

Article detail · 2015

Comparison of Na, K, Cl, Hb and Hct values measured by blood gas analyzer and laboratory auto-analyzer

Journal

TURKISH JOURNAL OF BIOCHEMISTRY-TURK BIYOKIMYA DERGISI

ISSN 0250-4685

The ISSN points to another catalog journal; the name is from the YÖKSİS record.

YÖKSİS OpenAlex Open access · gold SJR Q3 JCR Q4 Citations 4 Percentile 59.9% FWCI 0.18
Year
2015
Type
article

Data source split

  • YÖKSİS YÖKSİS article record
  • YÖKSİS venue TURKISH JOURNAL OF BIOCHEMISTRY-TURK BIYOKIMYA DERGISI
  • Catalog match (ISSN) Turkish Journal of Biochemistry
  • OpenAlex OpenAlex enrichment (abstract, citations, topics)

Abstract

OpenAlex · English

Abstract Objective: Measurements of electrolyte levels made by blood gas analyzers or routine methods based on membrane and selective electrodes are expected similar performance characteristics that do not show significant differences. It is clinically important that those measurements should give equivalent results and confirm the closeness to the absolute value. In this study, it was aimed to compare sodium (Na), potassium (K), chlorine (Cl), hemoglobin (Hb) and hematocrit (Hct) values measured with blood gas analyzer and standard automatic devices in the laboratory (lab) from venous blood samples. Methods: A prospective, randomized study was conducted in simultaneous venous samples from antecubital regions of patients admitted to emergency department for various reasons, were analyzed for Na, K, Cl, Hb and Hct values on blood gas analyzer and standard automatic devices in the lab. The obtained data were compared statistically. Results: A total of 100 patients were included in the study. A statistically significant correlation was found between laboratory and venous blood gas (vbg) Na, K, Cl, Hb and Hct values (p<0.001). The average difference between them was -7.9 mEq/L (r=0.720) for Na, 0.5 mEq/L (r=0.785) for K, -5.1mEq/L (r=0.790) for Cl, -0.14 grams/dL (r=0.757) for Hb and -% 1.54 (r=0.749) for Hct, respectively. Their correlation was formulated (Lab Na=vkg Nax0.621+46.081; Lab K=vkg Kx0.23+1.977; Lab Cl=vkg Clx0.847+11.050; Lab Hb=vkg Hbx0.583+5.015; Lab Hct=vkg Hctx0.565+15.024). Conclusion: In our study, it was detected that there is a statistically significant positive correlation between venous blood Na, K, Cl, Hb and Hct values of blood gas analyzer and standard automated devices in lab, and closeness of the desired performance is provided.

Topics

Citations

OpenAlex cited_by_count. Not a WoS or Scopus citation count; those sources have no separate column here.

4 citations

OpenAlex cited_by_count (cache / database)

Authors

  1. Nalan Kozaci
  2. Mehmet Oguzhan Ay
  3. RAMAZAN GÜVEN İSTANBUL ATLAS ÜNİVERSİTESİ
  4. Muhammed Ikbal Sasmaz
  5. Adevviye Karaca