Özet
A distinguishability measure is defined for magnetic resonance-electrical impedance tomography (MR-EIT) based on magnetic flux density measurements. This general definition is valid for 2D and 3D structures of any shape. As a specific case, a 2D cylindrical body with concentric inhomogeneity is considered and a bound of the distinguishability is analytically formulated. Distinguishabilities obtained with potential and magnetic flux density measurements are compared.
Konular
Atıflar
OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.
6atıfOpenAlex · cited_by_count (önbellek / veritabanı)
Yerel katalogda bu makaleye atıf yapan 4 yayın (OpenAlex referans eşleşmesi; tam dünya listesi değildir).
- 2012 Image Reconstruction in Magnetic Resonance Conductivity Tensor Imaging MRCTIAtıf 8 · OpenAlex
- 2007 Current Injection Optimization for Magnetic Resonance-Electrical Impedance Tomography (MREIT)Atıf 3 · OpenAlex
- 2016 Two alternatives for magnetic resonance electrical impedance tomography injected or induced currentAtıf 1 · OpenAlex
- 2009 Regional Image Reconstruction with Optimum Currents for MREIT — Evaluation on Shepp-Logan Conductivity PhantomAtıf 1 · OpenAlex