Comparison of 3 pupillometers for determining scotopic pupil diameter

ISSN1120-6721
YÖKSİS OpenAlex
Yıl2012
Atıf9OpenAlex
Atıf10Semantic Scholar · 1 etkili
Yüzdelik%57,7
FWCI0,21,00 = dünya ortalaması
Scopus (SJR)Q2
WoS (JCR)Q4

Veri kaynağı ayrımı

  • YÖKSİSYÖKSİS makale kaydı
  • YÖKSİS dergi adıEuropean Journal of Ophthalmology
  • Katalog eşleşmesi (ISSN)European Journal of Ophthalmology
  • OpenAlexOpenAlex zenginleştirmesi (özet, atıf, konular)
  • Semantic Scholaratıf sayısı (OpenAlex ile birleştirilmez)

Özet

OpenAlex İngilizce

PURPOSE: The pupil diameter under low ambient illumination is diagnostically valuable for refractive surgery. The aim of study was to compare the NeurOptics® Pupillometer, Sirius®, and Ocular Wavefront Analyser® in determining scotopic pupil diameter. MATERIALS AND METHODS: A total of 96 eyes of 48 subjects were included. The scotopic pupil size was measured with 3 instruments and the measurements repeated following instillations of 1% cyclopentolate. Agreement between the instruments was assessed. RESULTS: The mean measurement obtained by Sirius was significantly larger than Ocular Wavefront Analyser and NeurOptics. The Ocular Wavefront Analyser measured significantly smaller than the others. The mean cycloplegic pupillary diameters (7.73±0.70 mm with NeurOptics, 7.42±0.45 mm with Ocular Wavefront Analyser, and 8.06±0.76 mm with Sirius) were significantly different obtained by 3 instruments (p<0.001, for each; one-way analysis of variance and paired t tests). CONCLUSIONS: The differences between measured pupil diameters with or without cycloplegia obtained by the NeurOptics, Sirius, and Ocular Wavefront Analyser were significant and have unacceptable levels of disagreement. These results may not be related to illumination and accommodation only, but also to measurement algorithms and technical differences of the devices.

Konular

Atıflar

OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.

9atıfOpenAlex · cited_by_count (önbellek / veritabanı)

Yerel katalogda bu makaleye atıf yapan 1 yayın (OpenAlex referans eşleşmesi; tam dünya listesi değildir).

  1. 2015 Assessment of Pupil Diameters in Pseudoexfoliation Syndrome under Scotopic, Mesopic, Photopic and Dynamic Conditions Using Infrared Pupillometer (Assessment of Pupil Diameters in Pseudoexfoliation Syndrome)Atıf 9 · OpenAlex

Yazarlar

8
  1. ÇİĞDEM ALTAN 1
  2. VEDAT KAYA 2
  3. BERNA BAŞARIR İSTANBUL SAĞLIK VE TEKNOLOJİ ÜNİVERSİTESİ 3
  4. HACI UĞUR ÇELİK 4
  5. ENGİN AZMAN 5
  6. SERPİL AKAR 6
  7. AHMET DEMİROK 7
  8. ÖMER FARUK YILMAZ 8