In this study, a novel a carbon paste electrode using the salt form of polyaniline (pani)-silicon dioxide composite sensitive to glucose, was prepared. Glucose oxidase enzyme was immobilized to carbon paste electrode by cross-linking with glutaraldehyde. Determination of glucose was carried out by the oxidation of enzymatically produced H2O2 at 0.4 V vs. Ag/AgCl. The effects of pH and temperature were investigated and optimum parameters were found to be 9.0 and 60 °C, respectively. The linear working range of the electrode was 5.0 × 10-6 - 1.0 × 10-3 M, R2 = 0.997. The storage stability and operation stability of the enzyme electrode were also studied. The results showed that 98.4 % of the response current was retained after 16 activity assays. The prepared glucose biosensor retained 37 % of initial activity after 70 days when stored in 0.1 M phosphate buffer solution at 4 °C.
Konular
Atıflar
OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.
18atıfOpenAlex · cited_by_count (önbellek / veritabanı)
Yerel katalogda bu makaleye atıf yapan 22 yayın (OpenAlex referans eşleşmesi; tam dünya listesi değildir).
- 2024 Green Synthesis of Silver Nanoparticles and Designing A New Amperometric Biosentsor to Determine Glucose LevelsAtıf 41 · OpenAlex
- 2024 Green synthesis of silver nanoparticles and designing a new amperometric biosensor to determine glucose levelsAtıf 40 · OpenAlex
- 2024 Green synthesis of silver nanoparticles and designing a new amperometric biosensor to determine glucose levelsAtıf 32 · OpenAlex
- 2015 Development of an acetylcholinesterase–choline oxidase based biosensor for acetylcholine determinationAtıf 31 · OpenAlex
- 2014 An amperometric biosensor for glucose detection from glucose oxidase immobilized in polyaniline polyvinylsulfonate potassium ferricyanide filmAtıf 29 · OpenAlex
- 2017 Glucose biosensor based on immobilization of glucose oxidase on a carbon paste electrode modified with microsphere-attached L-glycineAtıf 28 · OpenAlex
- 2016 Development of an acetylcholinesterase choline oxidase based biosensor for acetylcholine determinationAtıf 28 · OpenAlex
- 2017 Glucose biosensor based on immobilizationof glucose oxidase on a carbon paste electrodemodified with microsphere-attached L-glycineAtıf 27 · OpenAlex
- 2017 Glucose biosensor based on immobilization of glucose oxidase on a carbon paste electrode modified with microsphere-attached L-glycineAtıf 27 · OpenAlex
- 2017 Glucose biosensor based on immobilization of glucose oxidase on a carbon paste electrode modified with microsphere attached glycine.Atıf 27 · OpenAlex