Comparison of Goldmann applanation tonometry rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes

Dergi INTERNATIONAL JOURNAL OF OPHTHALMOLOGY
YÖKSİS OpenAlex
Yıl2015
Atıf27OpenAlex
Atıf36Semantic Scholar · 2 etkili
Yüzdelik%82,6
FWCI1,741,00 = dünya ortalaması
Scopus (SJR)Q3
WoS (JCR)Q4

Veri kaynağı ayrımı

  • YÖKSİSYÖKSİS makale kaydı
  • YÖKSİS dergi adıINTERNATIONAL JOURNAL OF OPHTHALMOLOGY
  • OpenAlexOpenAlex zenginleştirmesi (özet, atıf, konular)
  • Semantic Scholaratıf sayısı (OpenAlex ile birleştirilmez)

Özet

OpenAlex İngilizce

AIM: To compare the intraocular pressure (IOP) measurements obtained with the rebound tonometry (RT), dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) in normal and glaucomatous eyes and investigate the effects of central corneal thickness (CCT) and corneal curvature (CC) on IOP measurements. METHODS: One hundred and twenty-four eyes of 124 subjects were enrolled in this cross-sectional study. Fifty-six of participants were healthy individuals and 68 of them were glaucomatous patients. IOP was measured on each subject always in the same order, ICare RT-Pascal DCT-GAT, after a minimum interval of 10min between measurements. CCT and CC were measured using a rotating Scheimpflug camera before the IOP measurements in all subjects. One way repeated measures ANOVA, Pearson correlation coefficient and regression analysis, and Bland-Altman analysis was used for the statistical assessment. RESULTS: Mean IOP for all enrolled eyes was 16.00±3.80 mm Hg for GAT, 16.99±4.91 mm Hg for RT, and 20.40±4.44 mm Hg for DCT. Mean differences between GAT and RT was -1.75±3.41 mm Hg in normal (P<0.001) and -0.37±3.00 mm Hg in glaucomatous eyes (P=0.563). Mean differences between GAT and DCT was -4.06±3.42 mm Hg in normal (P<0.001) and -4.67±3.12 mm Hg in glaucomatous eyes (P<0.001). GAT and RT were significantly positive correlated with CCT in normal (r=0.317, P=0.017 and r=0.576, P<0.001, respectively) and glaucomatous eyes (r=0.290, P=0.016 and r=0.351, P=0.003, respectively). DCT was also significantly positive correlated with CCT in normal eyes (r=0.424, P=0.001) but not in glaucomatous eyes (r=0.170, P=0.165). All tonometers were unaffected by CC. CONCLUSION: IOP measurements by RT and DCT were significantly higher than GAT. DCT has highest IOP measurements among these tonometers. RT was most influenced tonometer from CCT although all tonometers were significantly positive correlated with CCT except DCT in glaucomatous eyes. CC did not influence IOP measurements.

Konular

Atıflar

OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.

27atıfOpenAlex · cited_by_count (önbellek / veritabanı)

Yerel katalogda bu makaleye atıf yapan 10 yayın (OpenAlex referans eşleşmesi; tam dünya listesi değildir).

  1. 2025 Comparison of icare rebound tonometer, non-contact tonometer and goldmann applanation tonometer in intraocular pressure measurementAtıf 1 · OpenAlex
  2. 2023 Ocular pulse amplitude and visual field changes in patients diagnosed with aortic regurgitationAtıf 1 · OpenAlex
  3. 2023 Ocular pulse amplitude and visual field changes in patients diagnosed with aortic regurgitationAtıf 1 · OpenAlex
  4. 2017 Reply to comment by De Bernardo and Rosa on “Evaluation of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in keratoconus”Atıf 1 · OpenAlex
  5. 2023 Ocular pulse amplitude and visual field changes in patients diagnosed with aortic regurgitationAtıf 0 · OpenAlex
  6. 2023 Ocular pulse amplitude and visual field changes in patients diagnosed with aortic regurgitationAtıf 0 · OpenAlex
  7. 2023 Ocular pulse amplitude and visual field changes in patients diagnosed with aortic regurgitationAtıf 0 · OpenAlex
  8. 2020 Comment on: Effect of Topical Hypotensive Medications for Preventing Intraocular Pressure Increase After Cataract Surgery in Eyes With GlaucomaAtıf 0 · OpenAlex
  9. 2018 Intraocular pressure elevation after phacoemulsification in eyes with pseudoexfoliation syndromeAtıf 0 · OpenAlex
  10. 2018 Intraocular pressure elevation after phacoemulsification in eyes with pseudoexfoliation syndromeAtıf 0 · OpenAlex

Yazarlar

4
  1. FATİH ÖZCURA KÜTAHYA SAĞLIK BİLİMLERİ ÜNİVERSİTESİ 1
  2. NİLGÜN YILDIRIM 2
  3. AFSUN ŞAHİN 3
  4. ERTUĞRUL ÇOLAK 4