False Immunosuppressant Measurement by LC-MS/MS Method Due to Radiopaque Agents
Veri kaynağı ayrımı
- YÖKSİSYÖKSİS makale kaydı
- YÖKSİS dergi adıCOMBINATORIAL CHEMISTRY HIGH THROUGHPUT SCREENING
- Katalog eşleşmesi (ISSN)Combinatorial Chemistry and High Throughput Screening
- OpenAlexOpenAlex zenginleştirmesi (özet, atıf, konular)
- Semantic Scholaratıf sayısı (OpenAlex ile birleştirilmez)
Özet
BACKGROUND: Although liquid chromatography coupled with tandem mass spectrometry (LC-MS/MS) is preferred as a reliable method, some molecules in the blood matrix may lead to false positive or false negative results. False positive or negative results show the direction of the deviation rate from the target value. AIM: The aim of this study was to investigate the effect of seven different radiopaque agents on four immunosuppressant drugs. METHODS: Every agent coded with RM1 to RM7 was added to control materials containing tacrolimus, everolimus, sirolimus, and cyclosporine A drugs. Measurements were performed using an LC-MS/MS instrument. Bias values were calculated to detect the deviation rates. RESULTS: All RMs led to false negative results in the tacrolimus and cyclosporine A levels at a rate of -19.77% (95% CI, -27.16 to 12.52) to -44.45% (95% CI, -49.20 to -39.69). The smallest deviations were seen in the everolimus levels with the administration of RM6 (gadodiamide) and in the sirolimus levels with RM1 (gadobutrol) at the rates of 4.04% (95% CI, -11.36 to -3.17) and 2.11% (95% CI, -7.18 to 7.11), respectively. The most affected drug by RM4 (gadopentetate dimeglumine salt) was sirolimus at the rate of 114.01% (95% CI, 97.31 - 130.76). RM5 (gadodiamide) interfered cyclosporine A at the most. The highest deviations were observed with the administration of RM3 (iohexol) in the everolimus and sirolimus levels at the rates of 153.72% (95% CI, 142.44 to 164.78) and 171.41% (95% CI, 157.91 to 184.97), respectively. CONCLUSION: Radiopaque agents interfered the measurement of immunosuppressant drugs. Especially, everolimus and sirolimus levels were affected due to using iohexol. The choice of gadodiamide or ioversol is important to reduce the risk of interference for everolimus measurement. The blood samples should be obtained for measurement of drug levels before contrast-enhanced imaging.
Konular
Atıflar
OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.
6atıfOpenAlex · cited_by_count (önbellek / veritabanı)
Yerel katalogda bu makaleye atıf yapan 11 yayın (OpenAlex referans eşleşmesi; tam dünya listesi değildir).
- 2020 The Role of Interferences in the Increasing Incidence of Vitamin D DeficiencyAtıf 9 · OpenAlex
- 2020 The Role of Interferences in the Increasing Incidence of Vitamin D DeficiencyAtıf 9 · OpenAlex
- 2020 The Role of Interferences in the Increasing Incidence of Vitamin D DeficiencyAtıf 9 · OpenAlex
- 2020 The Role of Interferences in the Increasing Incidence of Vitamin D DeficiencyAtıf 9 · OpenAlex
- 2021 Effects of Broad Spectrum Antibiotics on Measurement of Immunosuppressant DrugsAtıf 3 · OpenAlex
- 2020 Effects of Broad Spectrum Antibiotics on Measurement of Immunosuppressant DrugsAtıf 3 · OpenAlex
- 2024 Analytical Interference with Contrast Agents in Genetic Test Result for ThrombophiliaAtıf 2 · OpenAlex
- 2025 Deneysel Bir Çalışma: Multivitaminler Takrolimus Düzeyinin, Parasetamol ve Trimetoprim Sülfametoksazol ise Siklosporin Düzeyinin Düşük Ölçülmesine Neden OluyorAtıf 0 · OpenAlex
- 2022 Experimental Investigation of Immunoglobulin and Complement Concentrations in Exposure to IVIG, HBIG, Rituximab, Tocilizumab, and BevacizumabAtıf 0 · OpenAlex
- 2022 In-Vitro Investigation of the Effect of Anti-Inflammatory and Analgesics on Measurement of Tumor MarkersAtıf 0 · OpenAlex