Makale detayı · 2015
Comparison of Energy Resolution of NaI(Tl) Scintillation Detectors Obtained by Analogue and Digital Ways
Radiation Science and Technology
- Yıl
- 2015
- ISSN
2575-5935- Tür
- article
Veri kaynağı ayrımı
- YÖKSİS YÖKSİS makale kaydı
- OpenAlex OpenAlex zenginleştirmesi (özet, atıf, konular)
Özet
İngilizce (OpenAlex)
Gamma-ray energy spectra were acquired using a Multichannel Analyzer (MCA) and a digitizer. A spectrometer which was consisted of a NaI(Tl) inorganic scintillation detector and a digitizer were used to obtain energy spectra of 137Cs radioisotope. The energy resolution values from analog and digital ways were calculated and compared with each other. The average analog and digital energy resolution values were determined as 7.69 % and 6.80 %, respectively. Obtained results showed that the digital systems are more preferable than the analog ones from the energy resolution point of view
Konular
- Radiation Detection and Scintillator Technologies
- Medical Imaging Techniques and Applications
- Nuclear Physics and Applications
Birincil konu Radiation Detection and Scintillator Technologies