Six SIGMA-based quality assessment of biochemical analytes: A comparative analysis of clinical laboratory improvement amendments 1988 and 2024 standards

ISSN1452-8258
YÖKSİS OpenAlex Açık erişim · diamond
Yıl2025
Atıf1OpenAlex
Yüzdelik%66,1
FWCI0,591,00 = dünya ortalaması
Scopus (SJR)Q3
WoS (JCR)Q4

Veri kaynağı ayrımı

  • YÖKSİSYÖKSİS makale kaydı
  • YÖKSİS dergi adıJournal of Medical Biochemistry
  • Katalog eşleşmesi (ISSN)Journal of Medical Biochemistry
  • OpenAlexOpenAlex zenginleştirmesi (özet, atıf, konular)
  • Semantic Scholaratıf sayısı (OpenAlex ile birleştirilmez)

Özet

OpenAlex İngilizce

Background: Laboratories have a responsibility to produce accurate and reliable results to ensure patient safety and meet quality standards. Within the Six SIGMA quality management framework, various approaches may emerge depending on the total allowable error (TAE) limits defined by different standards. This study aimed to compare analyte SIGMA levels using CLIA 1988 and CLIA 2024 criteria, determine appropriate quality control procedures based on Westgard multirules, and identify potential sources of error using the Quality Goal Index. Methods: SIGMA values for 17 biochemical analytes were calculated based on internal and external quality control data using the formula (TEa - bias)/CV The Quality Goal Index (QGI) was determined using the formula bias/(1.5xCV). All analytes were evaluated at the Karabuk Public Health Laboratory between November 2024 and March 2025. Results: Amylase (Levels 1 and 2), total bilirubin (Level 1), high-density lipoprotein cholesterol (Levels 1 and 2), creatine kinase (Levels 1 and 2), and lactate dehydrogenase (Levels 1 and 2) demonstrated world-class SIGMA performance according to both CLIA 1988 and CLIA 2024 standards. However, a decline in SIGMA values was observed when calculated using the more stringent CLIA 2024 limits. Conclusions: The comparison of CLIA 1988 and CLIA 2024 standards demonstrated that stricter TEa limits can significantly impact the performance of SIGMA. While some analytes maintained world-class performance, others exhibited a notable decline, necessitating enhanced quality control measures. These findings emphasise the need for laboratories to periodically reassess test performance in light of evolving regulatory standards to ensure continued analytical reliability and patient safety.

Konular

Atıflar

OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.

1atıfOpenAlex · cited_by_count (önbellek / veritabanı)

Yazarlar

2
  1. MERT ÜĞE MANİSA CELÂL BAYAR ÜNİVERSİTESİ 1
  2. SALİHA AKSUN 2