İçeriğe geç
akaturk Akademik ölçüm

Makale detayı · 2022

In Vitro Quantitative Evaluation of Postprocessing Filter for Metal Artifact Reduction in Cone Beam Computed Tomography Images of Titanium and Zirconium Dioxide Implants

Dergi

BioMed Research International

ISSN 2314-6133

YÖKSİS OpenAlex Açık erişim · hybrid SJR Q2 JCR Q3 Atıf 2 Yüzdelik 56.7% FWCI 0.4
Yıl
2022
Tür
article

Veri kaynağı ayrımı

  • YÖKSİS YÖKSİS makale kaydı
  • YÖKSİS dergi adı BIOMED RESEARCH INTERNATIONAL
  • Katalog eşleşmesi (ISSN) BioMed Research International
  • OpenAlex OpenAlex zenginleştirmesi (özet, atıf, konular)

Özet

OpenAlex · İngilizce

Objective: To evaluate a postprocessing filter of a new imaging-processing software for analysis of metal artifact reduction. Methods: Eight artificial edentulous mandibles (phantoms), where titanium and zirconium dioxide implants had been installed in four different regions (i.e., incisors, canine, premolars, and molars). CBCT volume was acquired, and then, four types of filters were applied to the images: BAR filter and Multi-CDT NR filter (e-Vol DX) and Sharpening Filters 1x and 2x (OnDemand). Artifact was assessed by measuring the standard deviation (SD) of the gray values of filtered and unfiltered images. The comparison between implant material, teeth, and filters was performed by using ANOVA, whereas multiple comparisons were performed by using Bonferroni's test. The level of significance adopted was 5%. Results: The results showing higher SD values, which suggests a worse image, were obtained with titanium implants compared to zirconium dioxide ones. With regard to the four filters used, it can be seen that the lowest SD values were obtained with BAR and Multi-CDT NR filters and the highest with Sharpening Filters 1x and 2x, with no statistical difference between them, except regarding the molar region in titanium implants. Conclusion: The highest SD values were seen in zirconium dioxide implants, mainly in the region of anterior teeth. The BAR filter was found to be the most effective as its SD value decreased significantly, indicating that the image quality was improved.

Konular

Atıflar

OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.

2 atıf

OpenAlex cited_by_count (önbellek / veritabanı)

Yazarlar

  1. KAAN ORHAN ANKARA ÜNİVERSİTESİ