Makale detayı · 2025
Comparison of CBCT and Panoramic Radiography Pediatric Doses in Three Different Age Groups
International journal of radiation research
- Yıl
- 2025
- ISSN
2345-4229- Tür
- article
Veri kaynağı ayrımı
- YÖKSİS YÖKSİS makale kaydı
- OpenAlex OpenAlex zenginleştirmesi (özet, atıf, konular)
Özet
İngilizce (OpenAlex)
Background: A wide variety of radiation dose values can occur in Cone BeamComputed Tomography (CBCT) devices with different field of view (FOV) sizes.Radiation dose of current CBCT devices have been considerably reduced.This study compared effective radiation doses and organ absorption doses obtained from panoramic and CBCT imaging at various FOV sizes in children aged five, ten, and fifteen years.Materials and Methods: To calculate the organ doses and effective doses, a dose calculation software PC-based Monte Carlo (PCXMC) 2.0 based on Monte-Carlo simulation was used for CBCT and panoramic exposures.Both absorbed and effective doses were calculated for the simulated phantoms of 5, 10 and 15 years old separately.Results: The organ doses of thyroid and salivary glands measured with 6*6 ECO Scan CBCT were found to be lower when compared with panoramic radiography organ doses.Panoramic radiography effective doses were lower than all other CBCT modes in all age groups.Conclusion: Although it is stated that this study does not have diagnostic data, it is thought that 6×6 ECO Scan Mode of Newtom CBCT can be chosen instead of panoramic radiography in children aged 5 years due to the dose differences.
Konular
- Dental Radiography and Imaging
- Sinusitis and nasal conditions
- Radiology practices and education
Birincil konu Dental Radiography and Imaging