Accuracy and reliability of 3D stereophotogrammetry: A comparison to direct anthropometry and 2D photogrammetry

Dergi

The Angle Orthodontist

ISSN 0003-3219

ISSN kaydı başka bir dergiye işaret ediyor; ad YÖKSİS kaydından.

YÖKSİS OpenAlex Açık erişim · diamond SJR Q1 JCR Q2 Atıf 183 Üst %10 Yüzdelik 97.0% FWCI 5.52
Yıl
2015
Tür
article

Veri kaynağı ayrımı

  • YÖKSİS YÖKSİS makale kaydı
  • YÖKSİS dergi adı The Angle Orthodontist
  • Katalog eşleşmesi (ISSN) Angle Orthodontist
  • OpenAlex OpenAlex zenginleştirmesi (özet, atıf, konular)

Özet

OpenAlex · İngilizce

OBJECTIVE: To evaluate the accuracy of three-dimensional (3D) stereophotogrammetry by comparing it with the direct anthropometry and digital photogrammetry methods. The reliability of 3D stereophotogrammetry was also examined. MATERIALS AND METHODS: Six profile and four frontal parameters were directly measured on the faces of 80 participants. The same measurements were repeated using two-dimensional (2D) photogrammetry and 3D stereophotogrammetry (3dMDflex System, 3dMD, Atlanta, Ga) to obtain images of the subjects. Another observer made the same measurements for images obtained with 3D stereophotogrammetry, and interobserver reproducibility was evaluated for 3D images. Both observers remeasured the 3D images 1 month later, and intraobserver reproducibility was evaluated. Statistical analysis was conducted using the paired samples t-test, intraclass correlation coefficient, and Bland-Altman limits of agreement. RESULTS: The highest mean difference was 0.30 mm between direct measurement and photogrammetry, 0.21 mm between direct measurement and 3D stereophotogrammetry, and 0.5 mm between photogrammetry and 3D stereophotogrammetry. The lowest agreement value was 0.965 in the Sn-Pro parameter between the photogrammetry and 3D stereophotogrammetry methods. Agreement between the two observers varied from 0.90 (Ch-Ch) to 0.99 (Sn-Me) in linear measurements. For intraobserver agreement, the highest difference between means was 0.33 for observer 1 and 1.42 mm for observer 2. CONCLUSIONS: Measurements obtained using 3D stereophotogrammetry indicate that it may be an accurate and reliable imaging method for use in orthodontics.

Konular

Atıflar

OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.

183 atıf

OpenAlex cited_by_count (önbellek / veritabanı)

Yazarlar

  1. FURKAN DİNDAROĞLU
  2. PINAR KUTLU
  3. GÖKHAN SERHAT DURAN ÇANAKKALE ONSEKİZ MART ÜNİVERSİTESİ
  4. SERKAN GÖRGÜLÜ
  5. ERHAN ASLAN