Corneal biomechanical properties of patients with pseudoexfoliation syndrome

Cornea

YÖKSİS OpenAlex SJR Q1 JCR Q2 Atıf 12 Yüzdelik 79.9% FWCI 1.21
Yıl
2011
ISSN
0277-3740
Tür
article

Veri kaynağı ayrımı

  • YÖKSİS YÖKSİS makale kaydı
  • OpenAlex OpenAlex zenginleştirmesi (özet, atıf, konular)

Özet

İngilizce (OpenAlex)

PURPOSE: To assess the corneal biomechanical properties of patients with pseudoexfoliation syndrome (PEX syndrome). METHODS: Fifty-two eyes of 52 patients (27 unilateral and 25 bilateral) and 42 eyes of 42 age-matched control subjects were enrolled in the study. Metrics of corneal biomechanical properties, including corneal hysteresis (CH) and corneal resistance factor (CRF), were measured with the ocular response analyzer (ORA). The ORA also determined the values of Goldmann-correlated intraocular pressure and corneal-compensated intraocular pressure. Central corneal thickness (CCT) was measured by the ORA integrated handheld ultrasonic pachymeter. The values recorded by the ORA were compared between eyes with PEX syndrome and those without PEX syndrome. Eyes with unilateral PEX syndrome and fellow eyes without PEX syndrome were also compared. RESULTS: The mean CH and CRF of all eyes with PEX syndrome were significantly lower than those of control eyes (P < 0.05). Neither corneal-compensated intraocular pressure nor Goldmann-correlated intraocular pressure showed a statistically significant difference in both the groups. Mean CCT values also did not show significant difference. Comparison of the unilateral eyes with PEX syndrome with the apparently normal fellow eyes revealed no significant difference between the mean CCT, mean CH, and mean CRF values (P > 0.05). However, the mean corneal-compensated intraocular pressure and Goldmann-correlated intraocular pressure were relatively higher in the PEX syndrome eye (P < 0.05). There was no significant difference between the mean CCT values of these groups (P > 0.05). CONCLUSIONS: The CH and CRF decrease in both unilateral and bilateral PEX syndrome suggested that PEX syndrome has a weakening effect on corneal biomechanical properties. Moreover, in subjects with clinically unilateral PEX syndrome, these properties were closer to eyes with PEX syndrome than to normal control eyes.

Konular

  • Corneal surgery and disorders
  • Glaucoma and retinal disorders
  • Ophthalmology and Visual Impairment Studies

Birincil konu Corneal surgery and disorders

Yazarlar

  1. Yenerel Nursal Melda
  2. Gorgun Ebru
  3. Kucumen Raciha Beril
  4. Oral Deniz
  5. Dinc Umut Aslı
  6. Ciftci Ferda
  7. RACİHA BERİL KÜÇÜMEN YEDİTEPE ÜNİVERSİTESİ