Safety evaluation of a new setup for transcranial electric stimulation during magnetic resonance imaging

ISSN1935-861X
YÖKSİS OpenAlex Açık erişim · gold
Yıl2021
Atıf15OpenAlex
Yüzdelik%76,7
FWCI1,151,00 = dünya ortalaması
Scopus (SJR)Q1
WoS (JCR)Q1

Veri kaynağı ayrımı

  • YÖKSİSYÖKSİS makale kaydı
  • YÖKSİS dergi adıBrain Stimulation
  • Katalog eşleşmesi (ISSN)Brain Stimulation
  • OpenAlexOpenAlex zenginleştirmesi (özet, atıf, konular)
  • Semantic Scholaratıf sayısı (OpenAlex ile birleştirilmez)

Özet

OpenAlex İngilizce

BACKGROUND: Transcranial electric stimulation during MR imaging can introduce safety issues due to coupling of the RF field with the stimulation electrodes and leads. OBJECTIVE: To optimize the stimulation setup for MR current density imaging (MRCDI) and increase maximum stimulation current, a new low-conductivity (σ = 29.4 S/m) lead wire is designed and tested. METHOD: The antenna effect was simulated to investigate the effect of lead conductivity. Subsequently, specific absorption rate (SAR) simulations for realistic lead configurations with low-conductivity leads and two electrode types were performed at 128 MHz and 298 MHz being the Larmor frequencies of protons at 3T and 7T. Temperature measurements were performed during MRI using high power deposition sequences to ensure that the electrodes comply with MRI temperature regulations. RESULTS: The antenna effect was found for copper leads at ¼ RF wavelength and could be reliably eliminated using low-conductivity leads. Realistic lead configurations increased the head SAR and the local head SAR at the electrodes only minimally. The highest temperatures were measured on the rings of center-surround electrodes, while circular electrodes showed little heating. No temperature increase above the safety limit of 39 °C was observed. CONCLUSION: Coupling to the RF field can be reliably prevented by low-conductivity leads, enabling cable paths optimal for MRCDI. Compared to commercial copper leads with safety resistors, the low-conductivity leads had lower total impedance, enabling the application of higher currents without changing stimulator design. Attention must be paid to electrode pads.

Konular

Atıflar

OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.

15atıfOpenAlex · cited_by_count (önbellek / veritabanı)

Yerel katalogda bu makaleye atıf yapan 3 yayın (OpenAlex referans eşleşmesi; tam dünya listesi değildir).

  1. 2021 Sensitivity and resolution improvement for in vivo magnetic resonance current‐density imaging of the human brainAtıf 11 · OpenAlex
  2. 2024 MR imaging of the magnetic fields induced by injected currents can guide improvements of individualized head volume conductor modelsAtıf 5 · OpenAlex
  3. 2023 Volumetric measurements of weak current–induced magnetic fields in the human brain at high resolutionAtıf 5 · OpenAlex

Yazarlar

6
  1. Frodi Gregersen 1
  2. CİHAN GÖKSU İZMİR YÜKSEK TEKNOLOJİ ENSTİTÜSÜ 2
  3. Gregor Schaefers 3
  4. Rong Xue 4
  5. Axel Thielscher 5
  6. Lars G Hanson 6