İçeriğe geç
akaturk Akademik ölçüm

Makale detayı · 2011

Kirschner wire fixation for scaphoid fractures an experimental study in synthetic bones

Journal of Hand Surgery-European Volume

YÖKSİS OpenAlex SJR Q1 JCR Q3 Atıf 11 Yüzdelik 78.2% FWCI 1.14
Yıl
2011
ISSN
1753-1934
Tür
article

Veri kaynağı ayrımı

  • YÖKSİS YÖKSİS makale kaydı
  • OpenAlex OpenAlex zenginleştirmesi (özet, atıf, konular)

Özet

İngilizce (OpenAlex)

We have studied the biomechanical stability in vitro of three different Kirschner (K) wire configurations in three types of simulated scaphoid waist fractures. The fractures were created with a saw in Sawbones models. There were three fracture patterns: perpendicular to the long axis of the scaphoid model; and 30° and 20° oblique to that. Two 1.2 mm. K-wires were used in each scaphoid. The three configurations were: parallel; 20° oblique; and crossing. The oblique or crossing configurations of K-wires were the most stable depending on the fracture pattern.

Konular

  • Orthopedic Surgery and Rehabilitation
  • Congenital limb and hand anomalies
  • Bone fractures and treatments

Birincil konu Orthopedic Surgery and Rehabilitation

Yazarlar

  1. ZAFER VOLKAN GÖKÇE
  2. HAKAN OFLAZ GEBZE TEKNİK ÜNİVERSİTESİ
  3. Dülgeroğlu Ali Murat
  4. Bora Arslan
  5. İZGE HAKAN GÜNAL