İçeriğe geç
akaturk Akademik ölçüm

Makale detayı · 2016

Rare Cell Immobilization on MEMS Scale Sensors Using Micro-Electromagnets

IEEE SENSORS JOURNAL

YÖKSİS OpenAlex SJR Q1 JCR Q1 Atıf 8 Yüzdelik 68.5% FWCI 0.61
Yıl
2016
ISSN
1530-437X
Tür
article

Veri kaynağı ayrımı

  • YÖKSİS YÖKSİS makale kaydı
  • OpenAlex OpenAlex zenginleştirmesi (özet, atıf, konular)

Özet

İngilizce (OpenAlex)

A novel method is presented for immobilizing rare single cells on MEMS scale sensor sites. The presented method does not require immobilization site to be chemically functionalized. Immobilizing single cells precisely on MEMS sensor sites is a requirement for many single cell analysis methods-reliability of cell analysis directly depends on the reliability of immobilization. The presented method uses magnetic forces to immobilize cells labeled with superparamagnetic beads on MEMS scale sites of interest. It can be coupled with many cell analysis techniques to form MEMS lab-on-a-chip devices due to the simple fabrication setup and principle of operation. The performance of the device was demonstrated with unbound superparamagnetic beads as well as with human breast tumor cells from the cell line MCF-7. Immobilization percentage and area were 95.2% and 1000 μm2, respectively, for superparamagnetic beads. MCF-7 cells labeled with superparamagnetic beads were immobilized at a percentage of 69.2% with the average distance from immobilized cells to micro-electromagnet being 50.8 μm. Thus, the presented method is a good candidate for integration with microfluidic sensing and analysis platforms.

Konular

  • Microfluidic and Bio-sensing Technologies
  • Microfluidic and Capillary Electrophoresis Applications
  • Nanopore and Nanochannel Transport Studies

Birincil konu Microfluidic and Bio-sensing Technologies

Yazarlar

  1. Rajapaksha WRL Gajasinghe
  2. Alperen Toprak
  3. Sukru Ufuk Senveli
  4. YAĞMUR AKIN YILDIRIM BARTIN ÜNİVERSİTESİ
  5. Michelle Jones
  6. Tan A. Ince
  7. Onur Tığlı