How Do Differences of Dental Implants’ Internal Connection Systems Affect Stress Distribution? A 3-Dimensional Finite Element Analysis

Dergi Balkan Journal of Dental Medicine
ISSN2335-0245
YÖKSİS OpenAlex Açık erişim · diamond
Yıl2019
Atıf3OpenAlex
Yüzdelik%71,6
FWCI0,711,00 = dünya ortalaması

Veri kaynağı ayrımı

  • YÖKSİSYÖKSİS makale kaydı
  • YÖKSİS dergi adıBalkan Journal of Dental Medicine
  • OpenAlexOpenAlex zenginleştirmesi (özet, atıf, konular)
  • Semantic Scholaratıf sayısı (OpenAlex ile birleştirilmez)

Özet

OpenAlex İngilizce

Background/Aim: A factor affecting the success rate of dental implants, which has been used successfully for many years, is the implant-abutment connection system. The purpose of this study was to evaluate the stress distribution of different implant-abutment connection systems under different forces. Material and Methods: This in vitro study included a finite element analysis. In the study, the cylindrical and screwed dental implants available in 3 different diameters from 4 different companies were categorized into 12 different models. Two different scenarios of force application were conducted on each model in this study. In the first scenario, 100 N force and 100 N moment were applied in a vertical direction onto a point considered as the center of each tooth. In the second scenario, a 100 N force and moment were applied at a 45˚ angle in an oblique direction. Results: As a result of the forces applied to dental implants of different diameters from different companies, octagon implant-abutment connection systems had less stress accumulation than hexagon implant-abutment connection systems. In addition, when stress accumulation ratios were evaluated according to the diameter of the implants used, it was observed that 3 mm diameter implants accumulated more stress in bone than 4 mm diameter implants; there was no significant difference between 4 mm diameter implants and 5 mm diameter implants. Conclusions: Implant-abutment connection system is important for the longevity of implants under the forces. Therefore, this factor should be considered during implant selection.

Konular

Atıflar

OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.

3atıfOpenAlex · cited_by_count (önbellek / veritabanı)

Yerel katalogda bu makaleye atıf yapan 1 yayın (OpenAlex referans eşleşmesi; tam dünya listesi değildir).

  1. 2021 Comparison of Stresses in Monoblock Tilted Implants and Conventional Angled Multiunit Abutment-Implant Connection Systems in the All-on-Four ProcedureAtıf 12 · OpenAlex

Yazarlar

4
  1. ÖZGE ÖZDAL ZİNCİR İSTANBUL ÜNİVERSİTESİ-CERRAHPAŞA 1
  2. GÖKAY KARAPINAR 2
  3. MERAL ÜNÜR 3
  4. AHMET BÜLENT KATİBOĞLU 4