Makale detayı · 2024 · article
Fabrication of a Portable Magnetic Microcantilever Using Fe40Ni38Mo4B18 Amorphous Ribbon and Its Application as a Humidity Sensor by Coating with TiO2 Nanotubes
Veri kaynağı ayrımı
- YÖKSİSYÖKSİS makale kaydı
- YÖKSİS dergi adıMagnetochemistry
- Katalog eşleşmesi (ISSN)Magnetochemistry
- OpenAlexOpenAlex zenginleştirmesi (özet, atıf, konular)
- Semantic Scholaratıf sayısı (OpenAlex ile birleştirilmez)
Özet
Microcantilevers (MCs) are highly sensitive sensors capable of detecting mass changes on the surface at the nanogram and even picogram scale. In this study, microcantilevers were fabricated for the first time using the Sodick AP250L Wire electrical discharge machining (EDM) from amorphous 2826MB (Fe40Ni38Mo4B18) ferromagnetic ribbons. This method is advantageous because it allows for the simultaneous production of a large number of microcantilevers, with about 100 MCs being produced in a single manufacturing process. Additionally, a straightforward and cost-effective measurement system was developed to measure the resonance frequency and frequency shift of the MC entirely through magnetic means, a technique not previously reported in the literature. To evaluate the performance of the MC, we employed it as a humidity sensor. For the TiO2-NT-coated MC, a frequency shift of approximately 202 Hz was observed when the humidity level changed from 5% to 95% relative humidity (RH).
Konular
Atıflar
OpenAlex cited_by_count. WoS veya Scopus atıf sayısı değildir; o kaynaklar için ayrı kolon yoktur.
2atıfOpenAlex · cited_by_count (önbellek / veritabanı)
Yerel katalogda bu makaleye atıf yapan 1 yayın (OpenAlex referans eşleşmesi; tam dünya listesi değildir).